Core Developed Assay Datasheet
Analyte: SAA
Assay Conditions: Serum samples require a 1:800 dilution.
Cross Reactivity
The antibody pair is highly specific and displays insignificant cross reactivity with the following core-developed assays:
TTR (prealbumin)
The antibody pair may cross react with other core-developed or commercially available assays.
Standard Curve Range: 1.37-1000ng/ml
Recovery in human serum: 83%
Sensitivity (Minimum Detectable Concentration): 20.8ng/ml
Intra-assay Precision: 4-8%
Inter-assay Precision: 12.8% (N=6)