Core Developed Assay Datasheet
Analyte: Bcl-2
Assay Conditions: Serum samples require a 1:5 dilution.
Cross Reactivity
The antibody pair is highly specific and displays insignificant cross reactivity with the following core-developed assays:
CD105, fPSA, NSE
The antibody pair may cross react with other core-developed or commercially available assays.
Standard Curve Range: 100-50000 pg/ml
Recovery in human serum: 120%
Sensitivity (Minimum Detectable Concentration): 100pg/ml
Intra-assay Precision: 3-4%
Typical Standard Curve: