Core Developed Assay Datasheet
Core Developed Assay Datasheet
Analyte: Angiostatin
Assay Conditions: Serum samples require a 1:300 dilution.
Cross Reactivity
The antibody pair is highly specific and displays insignificant cross reactivity with the following core-developed assays:
Thrombospondin, Endostatin
The antibody pair may cross react with other core-developed or commercially available assays.
Standard Curve Range: 137-100000pg/ml
Recovery in human serum: 56%
Sensitivity (Minimum Detectable Concentration): 56pg/ml
Intra-assay Precision: 2-5%